Ultra High-Performance Liquid Chromatography System LZ-UHPLC-A100 is a stacked, modular UHPLC that separates and measures the components of complex liquid samples. A linear motor quaternary pump drives the mobile phase up to 70 MPa, a autosampler loads samples, and a column compartment holds the temperature steady. A PDA detector and an ELSD detector read the eluting peaks. It offers 0.001 to 10 mL/min flow. Labs use it for pharmaceutical, food and chemical analysis.
| Specification | Value |
|---|---|
| Pump type | Quaternary, linear motor digital-driven |
| Flow range | 0.001–10 mL/min, in 0.001 mL/min steps |
| Maximum operating pressure | 70 MPa (10,150 psi) |
| Flow precision | ≤ 0.06% RSD |
| Flow accuracy | ≤ ±0.3% |
| Gradient range | 0–100% |
| Degassing chamber volume | 480 µl |
| Autosampler type | Type II |
| Tray capacity | 1 mL × 330; 2 mL × 192 (standard); 4 mL × 108; 10 mL × 48 |
| Injection volume | 0.1–120 µl |
| Injection precision | Full loop ≤ 0.25% RSD; partial quantitative loop ≤ 0.5% RSD |
| Carry-over | ≤ 0.005% (specified cleaning procedure) |
| Column compartment temperature range | 5–80 ℃ above room temperature |
| Column compartment stability / accuracy | ±0.1 ℃ / ±0.5 ℃ |
| Column capacity | Up to 4 × 30 cm columns; optional column switching valve |
| PDA diode array | 1024 diodes |
| PDA light source | Deuterium lamp + tungsten lamp |
| PDA wavelength range | 200–800 nm, adjustable in 1 nm steps |
| PDA wavelength accuracy | ±1 nm |
| PDA noise / drift | ±5 × 10⁻⁶ AU / 1 × 10⁻⁴ AU/h |
| PDA sampling frequency | Up to 100 Hz |
| PDA spectral bandwidth | 10 nm |
| PDA flow cell path length | 5 mm (10 mm optional) |
| ELSD light source | 650 nm laser, 30 mW maximum |
| ELSD atomizer temperature | 3–80 ℃ above room temperature, 1 ℃ steps |
| ELSD drift tube temperature | 3–130 ℃ above room temperature, 1 ℃ steps |
| ELSD drift tube stability | ±0.03 ℃ |
| ELSD gas flow | 0.0–5.0 L/min |
| ELSD noise / drift | ≤ 0.015 mV / ≤ 0.025 mV per 30 min |
| Software | Chromatography workstation, 24-bit signal acquisition, 1–4 detector signals |