Flame Photometer LZ-FLP-B100 is a compact and reliable laboratory instrument designed for the quantitative determination of potassium (K), sodium (Na), lithium (Li), and calcium (Ca), Barium (Ba) in liquid samples. It features a 7.0-inch color touchscreen with an intuitive interface for simple operation and clear result display. With automatic flame-out detection, gas cut-off protection, calibration curve management, USB connectivity, and optional built-in printing, it provides convenient and dependable elemental analysis for routine laboratory testing.
| Test Elements | K / Na / Li / Ca / Ba |
| Number of Channels | 5 |
| LOD (ppm) | K: ≤0.156 ppm Na: ≤0.184 ppm Li: ≤0.104 ppm Ca: ≤2.004 ppm Ba: ≤6.043 |
| Linear Error (ppm) | K: ≤0.195 Na: ≤0.69 Li: ≤0.146 Ca: ≤3.006 Ba: ≤9.064 |
| Test Range (ppm) | K: 0.15–100br/>Na: 0.018–100 Li: 0.1–30 Ca: 0.1–1000 Ba: 0.1-300 |
| Response Time | <8 s |
| Sample Uptake | <4 mL/min |
| Stability | ≤3%(15s), ≤15%(6min) |
| Reproducibility | 3% |
| Fuel | LPG |
| Oxidant | Air |
| Air Supply | External |
| Power Consumption | 30 W |
| Power Supply | AC 100–240 V, 50/60 Hz |
| Working Conditions | Temperature: 10–35°C; ≤85% RH |
| External Dimensions | 425 x 310 x 437 mm |
| Packing Dimensions | Main Unit: 634 x 364 x 337 mm Air Compressor: 350 x 185 x 345 mm |
| Gross Weight | Main Unit: 13 kg Air compressor: 5.3 kg |
Suitable for clinical laboratories, pharmaceutical testing, environmental analysis, food and beverage testing, agricultural laboratories, water-quality analysis, and general chemical research requiring quantitative determination of K, Na, Li, and Ca.