B2 Biological Safety Cabinet LZ-BSC-C210 delivers controlled airflow from the front opening across the work area, passing through ULPA 99.9995% downflow and HEPA 99.97% exhaust filtration before complete external discharge. The biological cabinet maintains stable operating conditions and expels all air without recirculation.

Specifications :
Air exhaust 100%
Internal Work Space 0.72 m2
Normal Size 1.2 m
Average Airflow Velocity
Inflow0.53 m/s
Down flow0.35 m/s
Airflow Volume
Inflow477 m³/h
Exhaust1220 m³/h
Filter Efficiency
HEPA Exhaust99.97%(0.3 µ)
ULPA Down flow99.9995%( 0.1 to 0.2 µ)
KI-Discus containment & Microbiological Testing
Product Protection Test 1~8×106≤5CFU
Cross-contamination Test 1~8×106≤2CFU
Noise
NSF/ANSI 49<62 dBA
EN 12469<59 dBA
Material
Work Zonestainless steel, type 304
Main Bodysteel with white oven-baked epoxy-polyester
Side Wallsstainless steel, type 304
Electrical Sliding Window Yes
RMS ≤3 μm
Power Consumption 855 W
Electrical Supply 220V/50 Hz, 110V/60 Hz
External Dimension(WxDxH) With Base Stand 1340×790×2200 mm
Internal Dimension(WxDxH) 1250×575×625 mm
Package Dimension 1425×945×1710 mm
Gross Weight 310 kg
Net Weight 250 kg
Features :

Maintains uniform laminar airflow for safe handling of samples

Stainless steel interior ensures quick cleaning and hygiene

Intuitive control panel for easy operation and system monitoring

Safety-grade glass with integrated UV access for secure observation

Sturdy and compact design for stable placement in laboratory environments

Applications :

Aseptic Sample Transfer
Transfer of cultures and reagents without exposure to the surrounding environment

Micropipetting Procedures
Accurate dispensing and handling of small liquid volumes during experiments

Culture Feeding Activities
Addition of nutrients and media to ongoing cultures under controlled conditions

Hazardous Reagent Handling
Working with reactive or toxic substances requiring contained airflow

Sterile Sample Preparation
Preparation of samples in controlled conditions before analysis or processing

Accessories :

Support stand: 560mm (22.0'') , 660mm (34.0'') or 760mm (29.9''), with castor
Ultraviolet Lamp Kits, Optional mobile UV lamps
Prefilters