A2 Biological Safety Cabinet LZ-BSC-A220

A2 Biological Safety Cabinet LZ-BSC-A220 creates a controlled workspace for microbiological handling under bio safety cabinet airflow standards. The inflow at 0.53 m/s and downflow at 0.35 m/s maintain consistent air movement, supporting stable conditions across routine and continuous laboratory operations.a

A2 Biological Safety Cabinet LZ-BSC-A220

A2 Biological Safety Cabinet LZ-BSC-A220 - Specifications

Air circulation 70%
Air exhaust 30%
Internal Work Space 0.54 m2
Normal Size 0.9 m
Average Airflow Velocity
Inflow0.53 m/s
Down flow0.35 m/s
Airflow Volume
Inflow363 m³/h
Down flow658 m³/h
Exhaust363 m³/h
ULPA Filter Efficiency
Inflow99.9995%(0.1 to 0.2 µ)
Down flow99.9995%(0.1 to 0.2 µ)
Biosafety Protection test
Product Protection Test 1~8×106≤5 CFU
Cross-contamination Test 1~8×106≤2 CFU
Noise
NSF/ANSI 49<60 dBA
EN 12469<57 dBA
Material
Work Zonestainless steel, type 304
Main Bodysteel with white oven-baked epoxy-polyester
Side Wallsstainless steel, type 304
Fluorescent Illumination 800 to 1200 Lux
Electrical Sliding Window Yes
RMS ≤2.3 μm
Power Consumption 361 W
Electrical Supply 220V/50 Hz, 110V/60 Hz
External Dimension(WxDxH) With Base Stand 1023×771×2255mm
Internal Dimension(WxDxH) 900×600×678 mm
Package Dimension 1125×945×1717 mm
Gross Weight 175 kg
Net Weight 120 kg

A2 Biological Safety Cabinet LZ-BSC-A220 - Key Features

A2 Biological Safety Cabinet LZ-BSC-A220 - Applications

Molecular Biology Labs
Workspace for handling nucleic acids and performing cell-based experiments

Clinical Testing Rooms
Processing blood, tissue, and patient samples in controlled airflow

Pharmaceutical Experiment Labs
Managing compound preparation and formulation activities safely

Pathology and Immunology Centers
Use in antibody studies, histology, and staining procedures

Environmental Research Units
Safe handling of chemicals and trace substances during laboratory studies

A2 Biological Safety Cabinet LZ-BSC-A220 - Accessories

Support stand: 560mm (22.0'') , 660mm (34.0'') or 760mm (29.9''), with castor
Ultraviolet Lamp Kits, Optional mobile UV lamps
Prefilters
Illumination Light
Disposable Active Carbon Postfilters

Request a Quote

Labozon Class II A2 Biological Safety Cabinet provides a controlled workspace for handling biological samples. Air is drawn into the cabinet, filtered, and recirculated within the work area, following biosafety cabinet practices, while a portion is exhausted to reduce contamination exposure. It is suitable for clinical, research, and pharmaceutical laboratories.

Class I Biological Safety Cabinet

Labozon Class I Biological Safety Cabinet provides a secure workspace for handling specimens and equipment while maintaining controlled environmental conditions. With aligned airflow and coordinated procedures, testing progresses systematically, and the lab safety cabinet ensures efficient workflows, smooth task execution, and well-organized analytical operations across the facility.

Class II B2 Biological Safety Cabinets

Labozon Class II B2 Biological Safety Cabinet uses a total exhaust system where filtered air is discharged outside without recirculation. Continuous inward airflow limits exposure while maintaining a controlled working environment. The biosafety cabinet type is widely used in labs where chemical vapours are present alongside biological materials.

2026 © All Rights Reserved by Labozon.com