Class II B2 Biological Safety Cabinets
Labozon Class II B2 Biological Safety Cabinet uses a total exhaust system where filtered air is discharged outside without recirculation. Continuous inward airflow limits exposure while maintaining a controlled working environment. The biosafety cabinet type is widely used in labs where chemical vapours are present alongside biological materials.
Class II B2 Biological Safety Cabinets Models
| Air Exhaust | 1 |
| Work Safety Height | 200 mm |
| Maximum Opening Height | 440 mm |
| Work Surface Height | 750 mm |
| Ascending Wind Speed | 0.53 ± 0.025 m / s |
| Maximum Opening | 480 mm |
| Assessed Space | Safety Height 200 mm (8 inch) |
| Inlet Velocity | 0.53 ± 0.025 m/s |
| Outflow Velocity | 0.33 ± 0.025 m/s |
| ULPA Filter | Two, 99.9995 % efficiency at 0.12 μm, filter life indicator |
| Maximum Opening | 480 mm |
| Assessed Space | Safety Height 200 mm (8 inch) |
| Inlet Velocity | 0.53 ± 0.025 m/s |
| Out flow Velocity | 0.33 ± 0.025 m/s |
| ULPA Filter | Two, 99.9995 % efficiency at 0.12 μm, filter life indicator |
| Air exhaust | 100% |
| Internal Work Space | 1.06 m2 |
| Normal Size | 1.8 m |
| Average Airflow Velocity | | Inflow | 0.53 m/s | | Down flow | 0.30 m/s |
|
| Airflow Volume | | Inflow | 706 m³/h | | Exhaust | 1805 m³/h |
|
| Air exhaust | 100% |
| Internal Work Space | 0.9 m2 |
| Normal Size | 1.5 m |
| Average Airflow Velocity | | Inflow | 0.53 m/s | | Down flow | 0.30 m/s |
|
| Airflow Volume | | Inflow | 592 m³/h | | Exhaust | 1515 m³/h |
|
| Air exhaust | 100% |
| Internal Work Space | 0.72 m2 |
| Normal Size | 1.2 m |
| Average Airflow Velocity | | Inflow | 0.53 m/s | | Down flow | 0.35 m/s |
|
| Airflow Volume | | Inflow | 477 m³/h | | Exhaust | 1220 m³/h |
|
| Air exhaust | 100% |
| Internal Work Space | 0.54 m2 |
| Normal Size | 0.9 m |
| Average Airflow Velocity | | Inflow | 0.53 m/s | | Down flow | 0.35 m/s |
|
| Airflow Volume | | Inflow | 363 m³/h | | Exhaust | 927 m³/h |
|